Sign In | Join Free | My chinaqualitylighting.com |
|
Categories | Limited Oxygen Index Tester |
---|---|
Brand Name: | GOLD |
Model Number: | GD-ISO4589 |
Certification: | ISO |
Place of Origin: | China |
MOQ: | 1 set |
Price: | Negotiable |
Payment Terms: | L/C, T/T |
Supply Ability: | 10 sets per month |
Delivery Time: | 25 working days |
Packaging Details: | Plywood case |
Product name: | Limited Oxygen Index Tester |
Standard: | ISO 4589-2, ASTM D2863 |
Oxygen concentration measurement range: | 0~100% |
Accuracy of oxygen concentration: | ± 0.1% |
Working pressure: | 0.1 Mpa |
Glass cylinder: | 500mm high, inner diameter is 80mm |
Company Info. |
Chongqing gold mechnical and electrical equipment Co., Ltd |
Verified Supplier |
View Contact Details |
Product List |
ISO 4589-2 Limited Oxygen Index Tester With Paramagnetic Oxygen Sensor
Introduction:
The Limited Oxygen Index Tester is designed to determine the
minimum oxygen concentration required to sustain combustion of
materials under controlled conditions. Utilizing an imported
paramagnetic oxygen sensor and high-precision mass flow
controllers, this analyzer ensures accurate and reliable results
for homogeneous solids, laminated materials, foams, films, and
textiles. It is ideal for laboratories and industries focused on
fire safety evaluation, compliant with ISO 4589-2, GB/T 2406-2009,
and other international standards.
Key Features:
Paramagnetic Oxygen Sensor: Lifetime durability, eliminating frequent replacements.
Auto-Adjustment System: PLC touchscreen controls oxygen concentration automatically.
Gas Mixing Chamber: Ensures uniform gas distribution for precise measurements.
Dual Gas Flow Meters: ±1% accuracy for nitrogen and oxygen flow control (1.0–15.0 L/min).
Safety Compliance: Built-in automatic calibration and safety circuits.
The airflow in the combustion tube is 40mm/s±2mm/s, and the concentration adjustment accuracy is ±0.1%;
The airflow at the top of the combustion tube is 90mm±10mm/s;
The portable igniter sprays downward 16±4mm, and the flame can be adjusted in the range of 10~30mm;
The measurement time can reach 5min, and the accuracy is ±0.1s;
Dual flow meter and pressure gauge device: accuracy ±1%;
Working pressure: 0.1MPa;
The glass tube is 500mm high and the inner diameter is 80mm.
Specifications:
Oxygen Range | 0–100% (±0.1% accuracy) |
Combustion Cylinder | 500mm height, 80mm diameter |
Flow Velocity | 40mm/s ±2mm/s in combustion chamber |
Dimensions | 690×480×670mm (portable desktop design) |
Power Supply | AC 220V, 50/60Hz |
![]() |